BGT-120
BGT-120
Descrição :
The Bag Integrity Tester produced by MedIntegrity, the BGT-120, is based on the pressure-drop test principle. The method follows ASTM F2095-01 and complies with GAMP and FDA 21 CFR Part 11. It includes authority management and electronic signature support, with full validation support, including IQ, OQ, PQ and SOP.
Drawing on years of experience in integrity testing and proven pressure-drop methods, the BGT-120 offers the following features
Drawing on years of experience in integrity testing and proven pressure-drop methods, the BGT-120 offers the following features
- Os registros históricos podem ser impressos, armazenados e consultados
- Portátil e fácil de conectar
- Operação simples, precisão, resultado de teste confiável
- Correspondência mais adaptável ao procedimento de teste de luvas atual, exibição de data e hora
Parâmetros
| Dimensões | 250*420*220 milímetros | |
| Pesagem | 11,7 kg | |
| Fonte de energia | CA 220 V 50 Hz | |
| Poder nominal | 120 W | |
| Duração do teste | 2-8 minutos | |
| Faixa de pressão | 0-5000pa (Outra faixa de pressão pode ser feita pelo cliente) | |
| Valor de resposta | ≤50pa | |
| Barulho | ≤50dB(A) | |
| Modo de trabalho | Trabalho contínuo | |
| Fonte de gás externa | Ar comprimido\Gás nitrogênio\Gás inerte | |
| Sensibilidade | 0,1pa | |
| Trilha de auditoria | ≥3 anos | |
| Registro histórico | ≥12000 conjuntos | |
| Programa | 80 | |
| Usuário | 80 | |
| Bateria | Padrão | |
| Sistemas | Máquina autônoma | |
Produtos relacionados